3H (AC004-J00) City Technology Hydrogen Sulfide Sensor

Available on order
For Bulk requirement, Kindly click on enquiry now button

 

 

Product Details: 

Product Description
Brand City Technology
Model Number 3H (AC004-J00)
Operating Principle 3-electrode electrochemical
Measurement Range 0-200 ppm H2S
Maximum Overload 1000 ppm H2S
Sensitivity* 0.37 ± 0.07 μA/ppm
Response Time (T90)* <35 Seconds
Baseline Offset (clean air)* -0.6 to +1.9 ppm H2S equivalent
Zero Shift (+20°C to +40°C) 2 ppm H2S equivalent
Resolution 0.25 ppm (when used with
recommended electronics)
Repeatability 1% of signal
Output Linearity Linear
Recommended Load Resistor 10 Ω
Bias Voltage Not Required
Weight 29 g
Cap Glass Filled Polypropylene
Body Glass Filled Polypropylene
Orientation Any
Operating Temperature Range -40°C to +50°C
Recommended Storage Temp 0°C to 20°C
Operating Pressure Range Atmospheric ± 10%
Pressure Coefficient 0.008 ± 0.002% signal/mbar
Operating Humidity Range 15 – 90% RH non-condensing
Long Term Sensitivity Drift* <2% signal loss/month
Expected Operating Life Two years in air
Storage Life 6 months in CTL container

Above specifications mentioned are for quick reference only. Actual technical details has to be obtained from below catalog.

Description

3H (AC004-J00) Hydrogen Sulfide Sensor

3H With side tag and PCB pin connections

Key Features & Benefits:

• Robust, 3-Series packaging
• Range of accessories available

Poisoning
CiTiceLs are designed for operation in a wide range of environments and harsh conditions. However, it is important that exposure to high concentrations of solvent vapours is avoided, both during storage, fitting into instruments and operation. When using sensors with printed circuit boards (PCBs), degreasing agents should be used before the sensor is fitted. Do not glue directly on or near the CiTiceL as the solvent may cause crazing of the plastic.

Cross Sensitivity Table
Whilst CiTiceLs are designed to be highly specific to the gas they are intended to measure, they will still respond
to some degree to various other gases. The table below is not exclusive and other gases not included in the table
may still cause a sensor to react.

IMPORTANT NOTE : The cross sensitivity data shown below does not form part of the
product specification and is supplied for guidance only. Values quoted are based on
tests conducted on a small number of sensors and any batch may show significant
variation. For the most accurate measurements, an instrument should be calibrated
using the gas under investigation.