3HYE(AE006-J30) City Technology Hydrogen Sensor

Available on order
For Bulk requirement, Kindly click on enquiry now button

 

 

Product Details: 

Product Description
Brand City Technology
Model Number 3HYE(AE006-J30)
Technology 3-electrode electrochemical
Measurement Range 0-10,000 ppm H2
Maximum Overload 20,000 ppm H2
Sensitivity ∗ 3 ± 1 nA/ppm
Resolution 10 ppm
Response Time (T90∗)  <70 Seconds
Baseline Offset (Clean air) + 2 to -150 ppm equivalent
Zero Shift (+20°C to +40°C) -150 ppm equivalent
Repeatability 2% of signal
Linearity Linear
Recommended Load Resistor 10 Ω
Bias voltage Not Required
Weight 22g
Orientation Any
Operating Temperature Range -20°C to +50°C
Recommended Storage Temp 0°C to 20°C
Operating Pressure Range  Atmospheric ± 10%
Pressure Coefficient 0.006% signal /mbar
Operating Humidity Range 15-90% RH non-considering
Long Term Sensitivity Drift* <2% of Signal loss/month
Expected Operating Life Two years in air
Storage Life 6 months in original container

Above specifications mentioned are for quick reference only. Actual technical details has to be obtained from below catalog.

Description

3HYE(AE006-J30)  City Technology Hydrogen Sensor

With Side tag and PCB Pin connections.

Key Features and Benefits :

  1. Robust, 3 Series Packaging
  2. Range of accessories available

Poisoning
CiTiceLs are designed for operation in a wide range of environments and harsh conditions. However, it is important
that exposure to high concentrations of solvent vapours is avoided, both during storage, fitting into instruments
and operation.
When using sensors with printed circuit boards (PCBs), degreasing agents should be used before the sensor is
fitted. Do not glue directly on or near the CiTiceL as the solvent may cause crazing of the plastic.

Cross Sensitivity Table
Whilst CiTiceLs are designed to be highly specific to the gas they are intended to measure, they will still respond
to some degree to various other gases. The table below is not exclusive and other gases not included in the table
may still cause a sensor to react.

SAFETY NOTE
This sensor is designed to be used in safety critical applications. To ensure that the sensor and/or instrument
in which it is used, are operating properly, it is a requirement that the function of the device is confirmed by
exposure to target gas (bump check) before each use of the sensor and/or instrument. Failure to carry out
such tests may jeopardize the safety of people and property.